Cryocoolers¶
The cryocooler is manufactured by CYRO Industries of America, Inc..
Close-circle cryostats are installed in <QM>2 allowing samples to be comprehensively studied in a wide range of temperatures from 12K to 500K. The cryocooler can operate both helium (12K to 300K) and nitrogen gas (300K to 500K ) using a single stream flow. In-house nitrogen and helium gas are used for cryocoolers. Just by switching the knob outside of the hutch, the gas supply can be altered.
The external tip-heater prevents frost and maintains a dry tip that helps to create an ice-free sample environment at low temperatures. A platinum temperature sensor is installed to control the low alarm setting temperature that triggers back the temperature controller to energize and maintain the ice-free environment.
The main component of the cryostat can be divided into five different components:
-
(i) source : the refrigerator or the cryostat act as a source to cool a reliable cryogenic temperature for a long continuous periods
-
(ii) compressor
-
(iii) PID temperature controller : Cryogenic temperature controlled Model 336 manufactured by Lake Shore is used. The controller’s zone tuning feature allows you to measure and control temperatures seamlessly from 300 mK to over 1,500 K by automatically switching temperature sensor inputs [1].
-
(iv) Power supply : 30 V digital and 24V DC are used
-
(v) Gas flow controller : The digital flow automatically control and displays the flow rate
Gas flow parameters
Helium gas: Normal flow rate is between 8 – 12 liters/min, typically 11 l/min
Nitrogen gas : Normal flow rate is between 8 – 12 liters/min, typically 9 l/min.
Safety
- Do not touch the cold helium and nitrogen gas stream. Exposure to the skin of the cold gas could result in severe injuries.
- The dewar system has pressure relief protection installed. Do not remove pressure reliefs or check valves.
Motors
Motor | Commands | Comments |
---|---|---|
Temperature controller | sampleT | Set the temperature |
To open the lakeshore¶
- Step 1:Open terminal
- Step 2: >
cd
- Step 3: >
/mnt/home/chess_id4b/.cs-studio
- Step 4: click okay
- Step 5: In the file you can find the recent window for lakeshore.opi
- Step 6: click lakeshore 2
- Change the Range : Rang3/High
Practical guide of the temperature switch between nitrogen and helium