Emergency proceduce if the temperature controller fails or Sudden power failure at beamline

Talk to Beamline Scientist or operators --- DONOT TRY THIS WITHOUT ASKING

If you suddenly saw the temperature controller is not showing the desired temperature

  • If the your setup is in nitrogen condition (300K - 80K) and suddenly temperature is dropiing drastically and you don't have controls
  • Switch the valve Nitrogen to Helium

Image title

  • press off in the compressor (it is inside the hutch)

Image title

  • Change the EPIC PVs range manually
    • Go to screen 3 of the beamline computer
    • Find the CS-Studio for temperature
    • Click the output 1
    • Go to range and select Range3/High for Input A
    • Click the output 2
    • Go to range and select Range3/High for Input B
    • Click the output 3
    • Go to range and select On for Input C
    • Click the output 4
    • Go to range and select On for Input D

Image title